日本精品视频国产亚洲精品精品精品_欧美日韩免费区一区二_欧美性爱在线观看第二区_午夜av在线激情_亚洲欧美综合伊人看片综合_亚洲天天综合网久久_男男羞羞视频网站国产_欧美视频二区中文字幕_日韩欧美成人色网_yellow中文字幕2024最新

歡迎進(jìn)入蘇州德計(jì)儀器儀表有限公司網(wǎng)站! 固話:0512-68790450 手機(jī):13915418480 返回首頁 | 聯(lián)系我們
大家都在搜: 固緯安規(guī)測(cè)試儀,固緯耐壓測(cè)試儀,固緯LCR測(cè)試儀
產(chǎn)品中心/ product center 您的位置:網(wǎng)站首頁 > 產(chǎn)品中心 > > 交直流電子負(fù)載 > PEL-2020 PEL-2030 PEL-2040 PEL-2041 可程式直流電子負(fù)載
產(chǎn)品分類
products
  1. 交直流電子負(fù)載
  2. 查看更多
可程式直流電子負(fù)載

可程式直流電子負(fù)載

產(chǎn)品型號(hào): PEL-2020 PEL-2030 PEL-2040 PEL-2041

所屬分類:交直流電子負(fù)載

產(chǎn)品時(shí)間:2024-05-08

簡要描述:PEL-2000可編程直流電子負(fù)載,此電子負(fù)載可以工作在定電流、定電壓、定電阻模式下,還可模擬各種動(dòng)態(tài)負(fù)載狀況,透過序列功能進(jìn)行負(fù)載電壓、電流、負(fù)載的上升及下降斜率、電流準(zhǔn)位、持續(xù)時(shí)間等參數(shù)來編輯負(fù)載波形,同時(shí)提供外部波形信號(hào)控制以及V&I監(jiān)測(cè)功能,測(cè)試參數(shù)及狀態(tài)的存儲(chǔ)可高達(dá)120組,系統(tǒng)可根據(jù)自動(dòng)測(cè)試需求,隨時(shí)由儀器內(nèi)部的儲(chǔ)存單元EEPROM中呼叫出來使用

詳細(xì)說明:

特點(diǎn)
-Sequence function to do high speed load simulations
-Flexible configuration with mainframes and plug-in modules
-Multiple independent load inputs up to 8 channels in a mainframe
-Sequence function to do high speed load simulations
-Parallel connection of inputs for increased load capacity
-Parallel connection of inputs for static and dynamic tests
-Multiple channels run for dynamic tests under one clock control
-Maximum rated power for low range high resolution CC tests
-Von setting and Von on/off for start-up Von tests
-Program mode to create work routines for repetitive tests
-OPP/OCP/OVP/OTP protections
-External channel control/monitoring via an analog control connector
-Multi-Interface USB Device / Host, RS-232C, and GPIB (Optional)
-Multi-mainframe link up to 5 mainframes in a system
 
PEL-2020(1002)
 
POWER
20W
100W
 
 
CURRENT
0~2A
0~20A
 
 
VOLTAGE
1~80V
 
MIN.OPERATING VOLTAGE(DC)
1.0V at 2A
1V at 20A
 
 
STATIC MODE
CONSTANT CURRENT MODE
 
Range
0~2A
0~20A
 
 
 
Resolution
0.5mA
5mA
 
 
 
Accuracy
±0.1%±0.1%F.S.
±0.1%±0.2%F.S.
 
 
CONSTANT RESISTANCE MODE
 
Range
0.075Ω~300Ω(100W/16V)
3.75Ω~15K(100W/80V)
 
 
Resolution
12bits
 
 
Accuracy
300Ω : ± 0.1 ± 0.2%
15KΩ: ± 0.01 ± 0.1%
 
CONSTANT VOLTAGE MODE
 
Range
1~80V
 
 
Resolution
20mV
 
 
Accuracy
±0.05% ± 0.1%F.S.
 
DYNAMIC MODE
CONSTANT CURRENT MODE
 
T1&T2
0.025mS ~ 10mS / Res : 1uS
1mS ~ 30S / Res : 1mS
 
 
Accuracy
1uS / 1mS ± 100ppm
 
 
Slew Rate(Accuracy 5%)
0.32 ~ 80mA/uS
3.2 ~ 800mA/uS
 
 
 
Slew Rate Resolution
0.32mA/uS
3.2mA/uS
 
 
 
Current
0~2A
0~20A
 
 
 
Current Resolution
0.5mA
5mA
 
 
 
Current Accuracy
±0.4% F.S.
 
MEASUREMENT
VOLTAGE READBACK
 
Range
0~16V
0~80V
 
 
 
Resolution
0.5mV
2.5mV
 
 
 
Accuracy
±0.05% ± 0.05% F.S.
 
 
 
Range
0~2A
0~20A
 
 
 
Resolution
0.0625mA
0.625mA
 
 
 
Accuracy
±0.1% ± 0.1% F.S.
 
PROTECTIVE
Over Power Protection(2%)
 
Range
1~102W
 
 
 
Resolution
0.5W
 
 
 
Accuracy
 
 
 
Over Current Protection(2%)
 
Range
0~20.4A
 
 
 
Resolution
0.05A
 
 
 
Accuracy
 
 
 
Over Voltage Protection(2%)
 
Range
1~81.6V
 
 
 
Resolution
0.2V
 
 
 
Accuracy
 
 
 
 
Over Temperature Protection
85℃
 
Rated Power Protection(2 %)
 
Value
110W
 
 
 
Accuracy
 
 
 
GENERAL
SHORT CIRCUIT
 
Current(CC)
2.2/2A
22/20A
 
 
 
Voltage(CV)
0V
0V
 
 
 
Resistance(CR)
3.75Ω
0.075Ω
 
 
INPUT RESISTANCE(LOAD OFF)
800KΩ(Typical)
 

 

 
PEL-2030(250W/30W)
POWER
30W
30W
250W
CURRENT
0~5A
0~4A
0~40A
VOLTAGE
1~80V
MIN.OPERATING VOLTAGE(DC)
1.0V at 5A
1.0V at 4A
1.0V at 40A
STATIC MODE
CONSTANT CURRENT MODE
 
Range
0~5A
0~4A
0~40A
 
Resolution
1.25mA
1mA
10mA
 
Accuracy
±0.1%±0.1%F.S
±0.1%±0.1%F.S
±0.1%±0.2%F.S
CONSTANT RESISTANCE MODE
 
Range
0.3Ω~1.2KΩ(30W/16V)
15Ω~60K(30W/80V)
0.0375Ω~150Ω(250W/16V)
1.875Ω~7.5K(250W/80V)
 
Resolution
12bits
 
Accuracy
12KΩ : ±0.1 ± 0.2%
60KΩ: ±0.01 ± 0.1%
150Ω : ±0.1 ± 0.2%
7.5KΩ: ±0.01 ± 0.1%
CONSTANT VOLTAGE MODE
 
Range
1~80V
 
Resolution
20mV
 
Accuracy
±0.05% ± 0.1%F.S.
DYNAMIC MODE
CONSTANT CURRENT MODE
 
T1&T2
0.025mS ~ 10mS / Res : 1uS
1mS ~ 30S / Res : 1mS
 
Accuracy
1uS / 1mS + 100ppm
 
Slew Rate(Accuracy 5%)
0.8 ~ 200mA/uS
0.64 ~ 160mA/uS
6.4 ~ 1600mA/uS
 
Slew Rate Resolution
0.8mA/uS
0.64mA/uS
6.4mA/uS
 
Current
0~5A
0~4A
0~40A
 
Current Resolution
1.25mA
1mA
10mA
 
Current Accuracy
±0.4% F.S.
MEASUREMENT
VOLTAGE READBACK
 
Range
0~16V
0~80V
0~16V
0~80V
 
Resolution
0.5mV
2.5mV
0.5mV
2.5mV
 
Accuracy
±0.05% ± 0.05% F.S.
CURRENT READBACK
 
Range
0~5A
0~4A
0~40A
 
Resolution
0.15625mA
0.125mA
1.25mA
 
Accuracy
±0.1% ± 0.1% F.S.
PROTECTIVE
Over Power Protection(2%)
 
Range
1~30.6W
1~255W
 
Resolution
0.15W
1.25W
 
Accuracy
 
 
Over Current Protection(2%)
 
Range
0~5.1A
0~40.8A
 
Resolution
0.0125A
0.1A
 
Accuracy
 
 
Over Voltage Protection(2%)
 
Range
1~81.6V
1~81.6V
 
Resolution
0.2V
0.2V
 
Accuracy
 
 
 
Over Temperature Protection
85℃
Rated Power Protection(2 %)
 
Value
33W
275W
 
Accuracy
 
 
GENERAL
SHORT CIRCUIT
 
Current(CC)
5.5/5A
4.4/4A
44/40A
 
Voltage(CV)
0V
0V
0V
 
Resistance(CR)
15Ω
0.3Ω
1.875Ω
0.0375Ω
INPUT RESISTANCE(LOAD OFF)
800KΩ(Typical)

 

 

 
PEL-2040
PEL-2041
POWER
35W
350W
35W
350W
CURRENT
0~7A
0~70A
0~1A
0~10A
VOLTAGE
1~80V
2.5~500V
MIN.OPERATING VOLTAGE(dc)
1.0V at 7A
1V at 70A
2.5V at 1A
2.5V at 10A
STATIC MODE
CONSTANT CURRENT MODE
 
Range
0~7A
0~70A
0~1A
0~10A
 
Resolution
1mA
10mA
0.2mA
2mA
 
Accuracy
±0.1%±0.1%F.S.
±0.1%±0.2%F.S.
±0.1%±0.1%F.S.
±0.1%±0.2%F.S.
CONSTANT RESISTANCE MODE
 
Range
0.025Ω~100Ω(350W/16V)
1.25Ω~5K(350W/80V)
1.25Ω~5KΩ(350W/125V)
50Ω~200K(350W/500V)
 
Resolution
12bits
12bits
 
Accuracy
100Ω : ±0.1 ± 0.2%
5KΩ: ±0.01 ± 0.1%
5KΩ : ±0.02 ± 0.2%
200KΩ: ±0.005 ± 0.1%
CONSTANT VOLTAGE MODE
 
Range
1~80V
2.5~500V
 
Resolution
20mV
100mV
 
Accuracy
±0.05% ± 0.1%F.S.
±0.05% ± 0.1%F.S.
DYNAMIC MODE
CONSTANT CURRENT MODE
 
T1&T2
0.025mS ~ 10mS / Res : 1uS
1mS ~ 30S / Res : 1mS
0.025mS ~ 10mS / Res : 1uS
1mS ~ 30S / Res : 1mS
 
Accuracy
1uS / 1mS ± 100ppm
1uS / 1mS ± 100ppm
 
Slew Rate(Accuracy 5%)
0.001 ~ 0.28A/uS
0.01 ~ 2.8A/uS
0.16 ~ 40mA/uS
1.6 ~ 400mA/uS
 
Slew Rate Resolution
0.001A/uS
0.01A/uS
0.16mA/uS
1.6mA/uS
 
Current
0~7A
0~70A
0~1A
0~10A
 
Current Resolution
1mA
10mA
0.2mA
2mA
 
Current Accuracy
±0.4% F.S.
±0.4% F.S.
MEASUREMENT
VOLTAGE READBACK
 
Range
0~16V
0~80V
0~125V
0~500V
 
Resolution
0.5mV
2.5mV
4mV
16mV
 
Accuracy
±0.05% ± 0.05% F.S.
±0.05% ± 0.05% F.S.
CURRENT READBACK
 
Range
0~7A
0~70A
0~1A
0~10A
 
Resolution
0.175mA
1.75mA
0.032mA
0.32mA
 
Accuracy
±0.1% ± 0.1% F.S.
±0.1% ± 0.1% F.S.
PROTECTIVE
Over Power Protection(2%)
 
Range
1~357W
1~357W
 
Resolution
1.75W
1.75W
 
Accuracy
 
 
Over Current Protection(2%)
 
Range
0~71.4A
0~10.2A
 
Resolution
0.175A
0.025A
 
Accuracy
 
 
Over Voltage Protection(2%)
 
Range
1~81.6V
1~510V
 
Resolution
0.2V
1.25V
 
Accuracy
 
 
 
Over Temperature Protection
≒85℃
≒85℃
Rated Power Protection(2 %)
 
Value
385W
385W
 
Accuracy
 
 
 
GENERAL
SHORT CIRCUIT
 
Current(CC)
≒7.7/7A
≒77/70A
≒1.1/1A
≒11/10A
 
Voltage(CV)
0V
0V
0V
0V
 
Resistance(CR)
≒1.25Ω
≒0.025Ω
≒50Ω
≒1.25Ω
INPUT RESISTANCE(LOAD OFF)
800KΩ(Typical)
800KΩ(Typical

PEL-2000可編程直流電子負(fù)載,此電子負(fù)載可以工作在定電流、定電壓、定電阻模式下,還可模擬各種動(dòng)態(tài)負(fù)載狀況,透過序列功能進(jìn)行負(fù)載電壓、電流、負(fù)載的上升及下降斜率、電流準(zhǔn)位、持續(xù)時(shí)間等參數(shù)來編輯負(fù)載波形,同時(shí)提供外部波形信號(hào)控制以及V&I監(jiān)測(cè)功能,測(cè)試參數(shù)及狀態(tài)的存儲(chǔ)可高達(dá)120組,系統(tǒng)可根據(jù)自動(dòng)測(cè)試需求,隨時(shí)由儀器內(nèi)部的儲(chǔ)存單元EEPROM中呼叫出來使用。

PEL-2000系列包含兩種主控含機(jī)框PEL-2002(可容納雙模組)與PEL-2004(可容納四模組),採用模組化設(shè)計(jì),協(xié)助使用者保留未來可能會(huì)運(yùn)用到更高功率需求的擴(kuò)充性,目前提供4款負(fù)載模組分別為:PEL-2020(200W雙通道負(fù)載模組100W*2 ; PEL-2030 (280W雙通道負(fù)載模組30W+ 250W) ; PEL-2040(350W單通道負(fù)載模組);PEL-2041(350W單通道負(fù)載模組)],以PEL-2004來代表zui多可以組成8個(gè)通道同步使用,功率範(fàn)圍200W~350W,電流範(fàn)圍0~70A,電壓範(fàn)圍1~500V,使用者可依據(jù)實(shí)際需求,選擇的負(fù)載模組及控制模組進(jìn)行組合或自組負(fù)載系統(tǒng),可同步執(zhí)行多組電子負(fù)載模組拉載的應(yīng)用,同時(shí)每個(gè)模組採用隔離與地浮接方式,避免造成短路回路 ; 現(xiàn)今綠能產(chǎn)業(yè)蓬勃發(fā)展,PEL-2000 系列符合各式消費(fèi)性電子及車用能源設(shè)備測(cè)試的需求,透過外部排線並聯(lián)控制功能,zui多可以同時(shí)控制五個(gè)主機(jī)框,並讓總體的總使用功率達(dá)到7kVA。

PEL-2000系列配備RS232、USB、GPIB(選配)接口、GO/NG接口,同時(shí)還配備了驅(qū)動(dòng)程序和針對(duì)LabView、VB、C、C++等相關(guān)測(cè)試軟體範(fàn)例。PEL-2000系列主要應(yīng)用於多通道或單通道輸出的AC/DC電源、DC/DC電源模組、充電器及電源類電子元件性能的測(cè)試,對(duì)設(shè)計(jì)研發(fā)、生產(chǎn)線測(cè)試及產(chǎn)品檢測(cè)作業(yè)上提供快速有效的測(cè)試方案,同時(shí)具有的OPP、OCP、OVP、OTP、低壓保護(hù)以及正反向極性保護(hù)功能,更大幅提高了產(chǎn)品的可靠性、是工程測(cè)試及自動(dòng)測(cè)試環(huán)境得以信賴的產(chǎn)品。

 
A. AUTOMATICALLY SEQUENCE FUNCTION
The Sequence function allows each channel to change its load sink according to a predefined sequence at a rate of up to 100 s per step. Each sequence is able to run concurrently, under the control of one clock. This is one of the most powerful features of the PEL-2000 Series as it is able to realistically simulate a multi-output power supply load. Under Dynamic mode, the load current or load resistance pulses between two preset levels at a pre-defined speed up to 25 s per step. This is often used as the standard test procedure to verify the response of a power supply to quick load changes.

The picture above is an example of a sequence used as a load profile for a single output switching power supply. A load profile is programmed to simulate the current draw of a power supply load. By using a current probe to acquire a current waveform, the PEL-2000 is able to evaluate the performance of a power supply based on the load sequence that is programmed. An oscilloscope is then used to display the result.
 
B. MULTIPLE OPERATING MODE
C.C. ( Constant Current )
In Constant Current mode, a load unit will sink the amount of current programmed. Regardless of the voltage change, the current will stay the same. Two selectable ranges are available for constant current mode: High (HI) and low (LO) range. Low range has a higher resolution for high precision measurements. If the current exceeds the low range, high range must be used. The PEL-2000 Series offers the maximum rated power for the low measurement range. This unique feature enables high-resolution measurements at high voltage only if the load current doesn't exceed the rated low range level. Both static and dynamic modes are available for CC operation. Static mode can be used for stability tests and Dynamic mode can be used to test transient load performance of a power supply. Go/NoGo is supported for both high and low ranges as well as Static and Dynamic modes.

C.V. ( Constant Voltage )
In Constant Voltage mode the load unit will sink current whilst keeping the voltage constant. Under CV mode, single channel load modules support two levels of voltage (A Value, B Value) with a preset cut-off current limit. An A / B button is available on the load module front panels to switch the Constant Voltage between A Value and B Value for power supply load regulation testing. Dual channel load modules only support A Value for CV mode. To accommodate the response of a power supply, response rates during voltage transitions can be set to fast or slow under static test mode.

C.V.+C.C. ( Constant Voltage + Constant Current )
When using CV mode, a current limit can be set for CV+CC mode. If the input voltage is greater than the preset CV limit, the channel will operate in CV mode if the input current is less than the preset CC limit. When the input current exceeds the current limit, the channel will operate in CC mode.

C.R. ( Constant Resistance )
In Constant Resistance mode the load units will linearly sink current and voltage to match a set resistance. Two different CR values and two different rising and falling slew rates can be set under CR Mode for single channel load modules. Like CC mode, Constant Resistance mode supports both dynamic and static loads. As with the other modes, Go/No Go is also supported.
 
C. PARALLEL DYNAMIC LOADING
All the load channels in a PEL-2000 mainframe can be connected in parallel to perform any combination of static or dynamic loading. Under Dynamic mode, the load current or load resistance pulses between two preset levels at a predefined speed of up to 25 s per step. When the channels are connected in parallel, dynamic tests are synchronously clocked. The ability to perform parallel dynamic loading gives you the flexibility to perform dynamic tests to highpower power supplies without the need for a dedicated high power electronic load.
 
D. FRAME LINK
The PEL-2000 Series allows multiple mainframes to be linked together with standard MIL 20-pin connectors to provide higher power load capacity. A maximum of 5 mainframes maximum, including one master and 4 slaves, can be chained together to give a 7kW load capacity for high current and high power applications
 
E. RUN PROGRAM
Under "Program" mode, up to 12 programs each with 10 panel setup memories, can be edited to create work routines for repetitive tests. After a program has completed, the results of all test steps, including the Go/NoGo judgments will be shown on the screen. Up to 12 programs with a total of up to 120 panel-setup memories can be chained together to be run one after the other.
 
F. PROTECTION MODES
The PEL-2000 Series includes a number of protection modes: Over Current Protection (OCP), Over Voltage Protection (OVP), Over Power Protection (OPP), Reverse Voltage Protection (RVP), and Under Voltage Protection (UVP). The protection modes are useful to protect both the load modules and the DUT(s). A buzzer can be set for when a protection setting has been tripped. When a protection mode has been tripped, the load unit will display an alarm and stop sinking current/voltage. When a load unit is operating in CR or CV mode, the unit may need Over Current Protection to prevent excessive current being sunk. Over Current Protection stops the load from sinking more current than its recommended limit and prevents the load from burn-out damage. Over Voltage Protection is used to limit the amount of voltage sunk. If the OVP trips, the PEL-Series load will stop sinking voltage. Over Power Protection is used when the input power exceeds the specifications of the load. When OPP is tripped, the power will cease to be sunk. Reverse Voltage Protection prevents reverse voltage damage to the PEL-2000 Series up to the specified rating. When Reverse Voltage Protection has been tripped, an alarm tone will sound until the reverse voltage is removed. Under Voltage Protection will turn off the load when the voltage drops below a set limit.
 
G. Von VOLTAGE SETTINGS
Von Voltage is the voltage limit at which the load will start to sink current. The Von Voltage setting is necessary for some kinds of power supply if a heavy load is connected during load on. There are two operation modes for Von Voltage: Latch ON and Latch OFF. When Von Latch is set to ON, the load starts to sink current when input voltage exceeds Von, and will continue to sink current even if the voltage drops below the Von Voltage later. When Von Latch is set to OFF, the load starts to sink current when input voltage exceeds Von, but will stop sinking current when the voltage drops below the Von voltage setting. This is to prevent the power supply from running into oscillation when the output voltage goes below certain level.
 
H. EXTERNAL CHANNEL CONTROL
A six-pin analog control connector is available on the rear panel to externally turn a load on/off and to externally monitor the load input status of each channel, including input current, input voltage, and CC current or CV settings. The Voltage Monitor Output (VMON) and Current Monitor Output (IMON) output the load input voltage and load input current as a percentage of rating current/voltage. Where 0 volts = 0% rating and 10 volts = 100% load input rating voltage or current. The voltage monitor output is across pins 1 & 3, and the current monitor output is across pins 1 & 2. Below shows the pin configuration of the voltage and current monitor outputs.
To turn a load on, an active low voltage (0-1V) must be applied across Load On (pin 5) and GND (pin 1), similarly an active high voltage (4-5V) must be applied to turn a load off. The Load On input must be TTL.



留言框

  • 產(chǎn)品:

  • 您的單位:

  • 您的姓名:

  • 聯(lián)系電話:

  • 常用郵箱:

  • 省份:

  • 詳細(xì)地址:

  • 補(bǔ)充說明:

  • 驗(yàn)證碼:

    請(qǐng)輸入計(jì)算結(jié)果(填寫阿拉伯?dāng)?shù)字),如:三加四=7

在線客服 二維碼

掃一掃,關(guān)注我們